《電子技術(shù)應(yīng)用》
您所在的位置:首頁 > 測試測量 > 設(shè)計應(yīng)用 > 改進LCR儀表測量穩(wěn)定性的相位裕度優(yōu)化方法及硬件實現(xiàn)
改進LCR儀表測量穩(wěn)定性的相位裕度優(yōu)化方法及硬件實現(xiàn)
電子技術(shù)應(yīng)用
楊學(xué)文,陳靈強,馬凱元
河北工程大學(xué) 信息與電氣工程學(xué)院
摘要: 為解決LCR表在測量容性阻抗時的自激問題,設(shè)計了一種增強測量穩(wěn)定性的LCR儀表。通過分析測量前端自平衡電橋電路的相位裕度,采用相位補償電路解決因相位滯后造成的自激問題。應(yīng)用直接數(shù)字頻率合成技術(shù)結(jié)合相位校準(zhǔn)電路設(shè)計,產(chǎn)生相位準(zhǔn)確的正交測量信號。使用開關(guān)鑒相式鎖相放大器處理微弱信號,經(jīng)16位模數(shù)轉(zhuǎn)換器采集后輸入到微處理器進行矢量信號合成與輸出顯示。相比專業(yè)儀器,所設(shè)計樣機對電阻測量相對誤差可達(dá)0.5%,對大電容測量相對誤差可達(dá)1%。
中圖分類號:TN98 文獻標(biāo)志碼:A DOI: 10.16157/j.issn.0258-7998.245400
中文引用格式: 楊學(xué)文,陳靈強,馬凱元. 改進LCR儀表測量穩(wěn)定性的相位裕度優(yōu)化方法及硬件實現(xiàn)[J]. 電子技術(shù)應(yīng)用,2024,50(12):70-76.
英文引用格式: Yang Xuewen,Chen Lingqiang,Ma Kaiyuan. Phase margin optimization method and hardware realization to improve the measuring stability of LCR instrument[J]. Application of Electronic Technique,2024,50(12):70-76.
Phase margin optimization method and hardware realization to improve the measuring stability of LCR instrument
Yang Xuewen,Chen Lingqiang,Ma Kaiyuan
School of Information and Electrical Engineering, Hebei University of Engineering
Abstract: To address the self-excitation issue in the measurement of capacitive impedance using LCR meters, an LCR instrument has been developed to enhance measurement stability. By analyzing the phase margin of the self-balancing bridge circuit at the measurement front-end, a phase compensation circuit has been employed to resolve the self-excitation problem caused by phase lag. Utilizing direct digital frequency synthesis combined with a phase calibration circuit design, quadrature measurement signals of accurate phase are generated. A switch-phase discriminator lock-in amplifier is used to process weak signals, which are sampled by a 16-bit analog-to-digital converter and input into the microprocessor for vector signal synthesis and output display. Compared to professional instruments, the prototype designed has achieved a relative error of up to 0.5% for resistance measurements and up to 1% for large capacitance measurements.
Key words : self-balancing bridge;phase margin;direct digital frequency synthesis;phase-locked amplifier

引言

測量阻抗有許多方法[1-3],有電橋法、諧振法、自平衡法、網(wǎng)絡(luò)分析法等,常見的LCR表多使用自平衡法進行阻抗測量,該法基于自平衡電橋電路,適合直流至幾兆赫茲的低頻段測量。

本文基于自平衡法設(shè)計了可以增強容性器件測量穩(wěn)定性的LCR儀表。通過分析前端輸入電路的相位裕度,使用相位補償電路解決了測量高容性器件時引起的自平衡電路自激問題。使用鎖相放大技術(shù)和直接數(shù)字頻率合成技術(shù)提高測量精度。實驗表明所設(shè)計樣機與專業(yè)儀器測量結(jié)果較匹配,且在高容性輸入時可保持穩(wěn)定測量。


本文詳細(xì)內(nèi)容請下載:

http://ihrv.cn/resource/share/2000006251


作者信息:

楊學(xué)文,陳靈強,馬凱元

(河北工程大學(xué) 信息與電氣工程學(xué)院,河北 邯鄲 056038)